IEC Acoustic Output Tables
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–119
CS Probe
Parameter
B
M
B+M
D
B+D
B+rD
B+rD
+D
M+cM
cM+D
P_ (MPa)
1.93
1.93
1.93
3.40
3.40
1.93
3.40
1.93
1.93
I
spta
(mW/cm
2
)
103.1
184.6
253.3
1603.1
1671.8
105.0
888.4
520.2
1603.1
I
p
(mm)
50.1
48.3
50.1
41.5
41.5
37.2
37.2
37.2
37.2
w
pb6
(||) (mm)
(
ă
)(mm)
2.1
2.5
2.2
2.5
2.1
2.5
1.5
3.0
1.5
3.0
1.8
3.5
1.8
3.5
1.8
3.5
1.8
3.5
prr (kHz)
–
0.86
0.86
1.71
1.71
0.28
0.28
1.48
1.48
srr (Hz)
144
–
144
–
144
10
10
–
–
Output Beam
Dim. (||) (mm)
(
ă
)(mm)
10.3
13.2
10.3
13.2
10.3
13.2
10.3
15.0
10.3
15.0
10.3
15.0
10.3
15.0
10.3
15.0
10.3
15.0
f
awf
(MHz)
3.60
3.56
3.60
3.30
3.30
2.53
2.53
2.53
2.53
APF (%)
(Acoustic Power
up Fraction)
na
na
na
na
na
na
na
na
na
AIF (%)
(Acoustic
Initialization
Fraction)
na
na
na
na
na
na
na
na
na
Maximum Power
(mW)
31.46
8.18
29.15
55.09
76.06
79.02
77.54
66.24
66.24
I
ob
(mW/cm
2
)
23.14
6.02
21.44
35.66
51.08
53.00
52.04
43.59
43.59
Power-up Mode
na
na
na
na
na
na
na
na
na
Initialization
Mode
na
na
na
na
na
na
na
na
na
Acoustic Output
Freeze
yes
yes
yes
yes
yes
yes
yes
yes
yes
I
tt
(mm)
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
I
ts
(mm)
contact
contact
contact
contact
contact
contact
contact
contact
contact
Table 6–151. CS IEC Acoustic Output Information