IEC Acoustic Output Tables
LOGIQ 500 Advanced Reference Manual
2276614–100 Rev. 0
6–140
S316 Probe
Parameter
B
M
B+M
D
B+D
B+rD
B+rD
+D
M+cM
cM+D
P_ (MPa)
2.15
2.15
2.15
2.05
2.15
2.15
2.15
2.15
2.15
I
spta
(mW/cm
2
)
92.8
148.7
210.6
851.8
913.7
257.2
585.5
436.3
436.3
I
p
(mm)
58.9
55.7
58.9
62.3
62.3
57.5
57.5
57.5
57.5
w
pb6
(||) (mm)
(
ă
)(mm)
2.0
3.7
1.7
4.5
2.0
3.7
2.3
3.1
2.3
3.1
1.9
3.6
1.9
3.6
1.9
3.6
1.9
3.6
prr (kHz)
–
0.89
0.89
4.53
4.53
5.20
5.20
5.20
5.20
srr (Hz)
112
–
112
–
112
39
39
–
–
Output Beam
Dim. (||) (mm)
(
ă
)(mm)
13.0
16.0
13.0
16.0
13.0
16.0
13.0
14.5
13.0
14.5
13.0
14.5
13.0
14.5
13.0
14.5
13.0
14.5
f
awf
(MHz)
3.42
3.36
3.42
3.00
3.00
2.98
2.98
2.98
2.98
APF (%)
(Acoustic Power
up Fraction)
na
na
na
na
na
na
na
na
na
AIF (%)
(Acoustic
Initialization
Fraction)
na
na
na
na
na
na
na
na
na
Maximum Power
(mW)
16.83
5.00
16.22
27.98
39.20
91.88
65.54
36.42
36.42
I
ob
(mW/cm
2
)
8.09
2.40
7.80
14.84
20.24
48.18
34.21
19.07
19.07
Power-up Mode
na
na
na
na
na
na
na
na
na
Initialization
Mode
na
na
na
na
na
na
na
na
na
Acoustic Output
Freeze
yes
yes
yes
yes
yes
yes
yes
yes
yes
I
tt
(mm)
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
I
ts
(mm)
contact
contact
contact
contact
contact
contact
contact
contact
contact
Table 6–167. S316 IEC Acoustic Output Information
Summary of Contents for Logiq 500