![background image](http://html.mh-extra.com/html/ge/logiq-e9/logiq-e9_service-manual_128639213.webp)
GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 11
7-3-5 Capturing Network Logs with Network Sniffer (cont’d)
15.)Select Displayed. This will save only the filtered values rather than the entire capture.
16.)Select SAVE.
If you perform Alt+D or Gather Logs, these sniffer logs will be included in the zip file.
17.)Exit the sniffer application by clicking on the X in the upper right corner.
18.)If you have already performed Save as, you can continue without saving and quit the program.
Make sure you have saved your data before exiting the program.
Figure 7-11 Select Capture Display
Figure 7-12 Save Capture Data Question