Chapter 6 Service Adjustments
6 - 3
6-2-2 Access to Adjustments (cont’d)
Wide Screen Monitors
:
Brightness and Contrast adjustments can be done via touchpanel encoders under Utilities screen
(
Figure 6-2 "Wide Screen Monitors Brightness and Contrast adjustments" on page 6-3
All Monitor settings have been optimized at the factory, so normally, there should be no need for any
further adjustments.
Under special lighting conditions, you may want to adjust the Monitor screen’s backlight intensity.
Factory default Intensity is: 100% (also depends on the Room Profile setting).
For complete information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter
16, or the appropriate LOGIQ E9 Release Notes. See:
Table 1-8 "Basic User Manual and Release Notes
per LOGIQ E9 console" on page 1-20
To obtain test patterns for B/W or color adjustments, press Utility on the Touch Panel and select Test
Patterns.