Chapter 4 Functional Checks
4-37
Section 4-4
Software Configuration Checks
Section 4-5
Peripheral Checks
Check that peripherals work as described below:
Table 4-9 Software Configuration Checks
Step
Task to do
Expected Result(s)
1.
Check Date and Time setting
Date and Time are correct
2.
Check that Location (Hospital Name) is correct
Location Name is correct
3.
Check Language settings
Desired Language is displayed
4.
Check assignment of Printer Keys
The default function for Print1-3 Keys is P1 (store image);
P2 (print); P3 (USB Quick Save). Print1-3 Keys can also
be assigned as desired by the customer
5.
Check that all of the customer’s options are set up correct All authorized functions are enabled
Table 4-10 Peripheral Checks
Step
Task to do
Expected Result(s)
1.
Press (
FREEZE
)
Stop image acquisition.
2.
Press (
P2
) on the Control Panel
The image displayed on the screen is printed on B&W printer.
3.
Connect with Foot Switch on USB port and
press once.
To start image acquisition (the same function as (
FREEZE
) key).