GE
PROPRIETARY TO GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-93
Section 8-14
Loading Software
LOGIQ P3
- FRU part numbers for software.
1.) 5308912-6 LOGIQ P3 Expert Software - R1.0.5
2.) 5315327-6 LOGIQ P3 Pro Software - R1.0.5
3.) 5315328-6 LOGIQ P3 BW Software - R1.0.5
LOGIQ P3 BT10
- FRU part numbers for software.
1.) 5368259-6 LOGIQ P3 BT Expert software - R2.0.5
2.) 5397504-6 LOGIQ P3 BT Pro software - R2.0.5
3.) 5397506-6 LOGIQ P3 BT BW software - R2.0.5
4.) 5443046-4 LOGIQ P3 BT 3 Port BW Software - R2.0.4
NOTE:
While it is believed to be unnecessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE:
To Enter BIOS Press F2 (Arrow) Key immediately after power on and enter BIOS Password for
LP3 as lakshya.
NOTE:
Please ensure AC adapter is connected during system upgrade!
1.) Before loading the Software Backup the database for reference see
and
after loading the Software Restore the bacup database.
2.) Insert the disk labeled “System & Application Software” into the DVD drive.
3.) Properly turn off the scanner by momentarily pressing the
Power On/Off
Switch. Select
“Shutdown” from the System Exit menu.
4.) If the system will not shutdown normally, hold down the
Power On/Off
Switch until the light
turns from green to amber.
5.) Turn on the scanner. System will detect the DVD-RW automatically.
Figure 8-93 Shutdown Dialog Box