Chapter 3 Installation
3-9
Section 3-4
Preparing for Installation
3-4-1
Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered or damaged.
3-4-2
Physical Inspection
3-4-2-1
System Voltage Settings
•
Verify that the scanner is set to the correct voltage. The Voltage ratings for the LOGIQ P3 Scanner
is found on Rating label near the Circuit Breaker at the rear of the system
3-4-3
EMI Protection
This Unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
3-4-4
Checking the Components
When a new system arrives, check that none of the components are damaged or in short supply. If
shipping damage or shortage occurs, contact the address shown in Chapter 1.
WARNING
WARNING
Connecting a LOGIQ P3 scanner to the wrong voltage level will most likely destroy the
scanner.
CAUTION
C
If the Power Plug is Modified or replaced to Suit the local Conditions and regulations, Ground
continuity check should be performed between Ground Pin on the Plug and the Metal Part on
the LOGIQ P3. Refer Section 10-7 for more details.