GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
4 - 6
Section 4-2 - General Procedures
VSIO (ECG)
Leakage Current measured at (record the value) and meets
allowable limits. Equipment passed all required checks and is
ready for use.
Active ECG:
Utility
>
Applications
>
Settings
>
ECG
>
Show ECG Tools
.
Connect cables. Press Display ECG Tab on
TP. Verify ECG signal displayed on monitor.
Plug in the ECG cables with nothing attached
so they pick up noise.
Diagnostics
Battery Pack/
Battery Pack functional checks
VNAV
Volume Navigation with inside probe functional
Table 4-2 General System Functional Checks (Continued)
Section
Functional Check
Debrief Script