6-3-4 Main monitor adjustments (continued)
Monitor arm tilt adjustment
If the monitor is too loose or too tight when it is tilted, perform a tilt adjustment.
Table 6-6 Monitor Arm Tilt Adjustment
Step
Corresponding Graphic
1.
Tilt the monitor down and remove the
monitor rear cover.
If you have a thick wrench, it is easier to
reach the nuts by using the boxed end of the
wrench.
2.
If the monitor is too loose, turn the
adjustment nut (
A
) towards the back and
nut (
B
) towards the front to increase the
friction, using a 13 mm wrench.
Turn the nuts the opposite direction to
decrease the friction. The monitor should
stay in the position it is released at.
NOTE
: Be sure to tighten or loosen both
sides equally. Start with 1/4 turn and test full
90 degree movement before tightening or
loosening more.
3.
Reinstall the monitor rear cover.