Chapter 7 Diagnostics/Troubleshooting
7 - 75
Section 7-10
Setting Troubleshooting
Table 7-27 Setting Troubleshooting
Symptom
Solution
Incorrect group assignment of CMFN group in Breast Comment
Library page 2
The cm from nipple annotations are assigned to the same
group as the o'clock annotations. This group needs to be
adjusted in the utility pages for these annotations.
Assign as Group 5
Region presets have not been set up, If the European region is
used the finger and hip models are not functional.
Set the region to None.
Imaging preset selection is not highlighted on Imaging Preset
Manager
The issue is Utility > Imaging Preset Manager > Imaging
Preset Selections list items are not highlighted when selected.
All functions work, the selection is just not highlighted.
Print button with both Copy to Dataflow and USB QuickSave
does not save image to clipboard.
If both Copy to Dataflow and USB QuickSave are programmed
to the same print button, images may fail to save.
Do not program Copy to Dataflow and USB QuickSave to the
same print button.
GSDF is defaulted to on which may cause undesirable image
quality on PACS systems that do not support GSDF.
Change the default GSDF setting to off.
System crashes if multiple commas are placed in annotations. Avoid putting multiple periods, commas, etc. in annotations as
this will cause a system crash.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......