Basic Service Manual
8 - 22
Section 8-4 - Loading / Reloading / Upgrading the Software
Loading the base image and application software - full load (continued)
15.)Configure any new features (for example, Device Management) if applicable. Refer to the LOGIQ
Totus Basic User Manual for more information. For software options (for example, Breast Assistant,
Powered by Koios DS and Tricefy), restore the information you recorded before loading the full
ghost.
16.)Store the software thumb drive on the inside pocket of the left side cover. Reinstall any covers and
Verify that all screws and covers removed earlier have been installed.
Functional checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-7 Loading Software all Partitions Functional Checks
: Section
Functional Check
Debrief Script
Equipment passed all required checks and is ready for
use.
Probe/connector functional checks
System CFM and PWD functional checks