GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
8 - 50
Section 8-5 - Replacing Cables
8-5-8
Replacement of the UBE To OPIO Touch Panel HDMI Cable
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-5-8-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the OPIO Top Cover Set-T as described in
8-7-2 "Replacement of the OPIO Top Cover
3.) Remove the UOPIO-T Assy as described in
8-7-10 "Replacement of the UOPIO-T Assy" on page
4.) Remove the Elevation Arm Cover Set-U as described in
8-8-20 "Replacement of the Elevation Arm
5.) Remove the Top Cover Assy-T as described in
8-8-17 "Replacement of the Top Cover Assy-T" on
6.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
Table 8-26 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 30 minutes
Phillips screwdriver
Nipper
Cut resistant gloves
Knee Pad