GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-5 - Replacing Cables
8 - 51
8-5-8-2
Removal Procedure
Steps
Corresponding Graphic
1.
Unlock Cable Clamp.
2.
Unscrew 2 screws to remove the Cable Clamp.
3.
Cut 2 Cable Ties.
4.
Pull UBE To OPIO Touch Panel HDMI Cable to
remove from the T2 OPIO BASE.
5.
Unscrew 4 screws to remove the Unity Cable Cover
Brkt