GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-5 - Replacing Cables
8 - 67
8-5-12 Replacement of the ACDC Link Cable To UBE Cable
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-5-12-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
3.) Remove the Unity Base Tray as described in
8-8-2 "Replacement of the Unity Base Tray" on page
4.) Remove the Front Cover Assy-T as described in
8-8-3 "Replacement of the Front Cover Assy-T" on
5.) Remove the Front Cover Bottom Assy-T as described in
8-8-6 "Replacement of the Front Cover
Table 8-30 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad