GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
8 - 86
Section 8-6 - Replacing Monitor and Monitor Arm
8-6-7
Replacement of the HDU 23.8 inch Monitor LOGIQ Exx
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-6-7-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
8-6-7-2
Removal Procedure
8-6-7-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
Table 8-37 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Remove the Monitor Cable Cover.
2.
Unscrew 4 screws. Disconnect the UBE To Monitor
HDMI Cable, OPIO HUB To Monitor USB 2.0 Cable
and OPIO HUB To Monitor Power Cable.
3.
Unscrew 4 screws to remove the HDU 23.8 inch
Monitor LOGIQ Exx from the console.