Basic Service Manual
8 - 116
Section 8-7 - Replacing OPIO
8-7-22 Replacement of the Touch Panel Joycoder Set-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-7-22-1
Preparations
1.) Remove the OPIO Top Cover Set-T as described in
8-7-2 "Replacement of the OPIO Top Cover
2.) Remove the UOPIO-T Assy as described in
8-7-10 "Replacement of the UOPIO-T Assy" on page
3.) Remove the OPIO Laser Color Trackball with Cable as described in
Trackball assy-U_FLT20UFA-GJ" on page 8-101
4.) Remove OPIO Knob Set-T Assy as described in
8-7-16 "Replacement of the OPIO Knob Set-T" on
5.) Remove OPIO Lower Cover Set as described in
8-7-17 "Replacement of the OPIO Lower Cover
6.) Remove OPIO Button Set-T as described in
8-7-18 "Replacement of the OPIO Button Set-T and
the User Define Key Cap Set-T" on page 8-109
7.) Remove T2 OPIO Key Rubber as described in
8-7-19 "Replacement of the T2 OPIO Key Rubber"
Table 8-58 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Philips Screwdriver
Spanner 24
Cut resistant gloves
Knee Pad
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......