Basic Service Manual
Section 8-7 - Replacing OPIO
8 - 123
8-7-25-2
Removal Procedure
8-7-25-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Steps
Corresponding Graphic
1.
Disconnect the OPIO HUB To Speaker from the
console.
2.
Unscrew 3 screws to remove the Unity Speaker Assy
L from the console.
3.
Unscrew 3 screws to remove the Unity Speaker Assy
R from the console