Basic Service Manual
8 - 124
Section 8-7 - Replacing OPIO
8-7-26 Replacement of the UHUB-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-7-26-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the OPIO Top Cover Set-T as described in
8-7-2 "Replacement of the OPIO Top Cover
3.) Remove the UOPIO-T Assy as described in
8-7-10 "Replacement of the UOPIO-T Assy" on page
8-7-26-2
Removal Procedure
8-7-26-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
NOTE:
The UBE To OPIO HUB USB 3.0 Cable can be inserted to any of two USB ports (OPIO HUB).
Table 8-62 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Unscrew 2 screws and unlock the Cable Clamp.
2.
Disconnect cables which are connected to the
UHUB-T.
3.
Unscrew the 12 screws to remove the UHUB-T from
the console.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......