Basic Service Manual
8 - 132
Section 8-8 - Replacing Main Console
8-8-4
Replacement of the Front Dummy Cover Set-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-4-1
Preparations
1.) Remove the Unity Base Tray as described in
8-8-2 "Replacement of the Unity Base Tray" on page
2.) Remove the Front Cover Assy-T as described in
8-8-3 "Replacement of the Front Cover Assy-T" on
8-8-4-2
Removal Procedure
8-8-4-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”
Table 8-66 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
No tools required
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Push the T2 CWD Dummy Cover and T2 ECG
Dummy Cover from the back of the Front Cover
Assy-T.