Basic Service Manual
8 - 144
Section 8-8 - Replacing Main Console
8-8-16 Replacement of the Rear Cover Bottom Assy-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-16-1
Preparations
Remove the Rear Cover Mid Assy-T as described in
8-8-10 "Replacement of the Rear Cover Mid Assy-
8-8-16-2
Removal Procedure
8-8-16-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”
Table 8-78 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
No tools required
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1. Pull the Rear Cover Bottom Assy-T to remove from
the console.