Section 8-8 - Replacing Main Console
8 - 147
8-8-19 Replacement of the T2 Handle
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-19-1
Preparations
1.) Remove the T2 Side Cover L Assy as described in
8-8-7 "Replacement of the T2 Side Cover L
2.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
3.) Remove the Rear Cover Mid Assy-T as described in
8-8-10 "Replacement of the Rear Cover Mid
4.) Remove the Rear Cover Top Assy-U as described in
8-8-15 "Replacement of the Rear Cover Top
8-8-19-2
Removal Procedure
8-8-19-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-81 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 10 minutes
Phillips Screwdriver
Hexagonal (Allen) Wrench
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1. Unscrew 4 screws to remove the Unity Cable Cover
Brkt from the console.
2. Unscrew 6 screws to remove the T2 Handle from the
console
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......