Basic Service Manual
8 - 162
Section 8-9 - Replacing Front End Parts
8-9-2
Replacement of the UAC-T PWA
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-9-2-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove Unity Base Tray as described in
8-8-2 "Replacement of the Unity Base Tray" on page 8-
3.) Remove Front Cover Assy-T as described in
8-8-3 "Replacement of the Front Cover Assy-T" on
4.) Remove Unity Protector Assy as described in
8-8-5 "Replacement of the Unity Protector Assy" on
8-9-2-2
Removal Procedure
8-9-2-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-89 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Unscrew 8 screws to remove the UAC&UCO-T Assy.
2.
Unscrew 9 screws to remove the UAC-T and Pull the
UAC-T PWA to remove the UAC-T PWA from the
UCO-T&UAC Assy.
Note: If the CWC-T is installed, remove the CWC-T as
described in
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......