Basic Service Manual
Section 8-11 - Replacing Options
8 - 203
8-11-15-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
4
.
Unscrew 2 screws and remove the BW Printer Assy
from the console.
5
.
Rotate the UBE To BW Printer Power Cable
counterclockwise and disconnect from the BW
Printer. Disconnect the UBE To BW Printer USB
Cable to remove from the BW printer.
Steps
Corresponding Graphic