Basic Service Manual
8 - 212
Section 8-11 - Replacing Options
8-11-19-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
6.
Disconnect the UVNAV PWR Cable from the VNAV
Assy.
7.
Cut the cable tie.
8.
Unscrew 3 screws to remove the T2 VNAV FIX BRKT
from the VNAV.
9.
Disconnect the UCO To V-NAV Signal Cable to the
VNAV.
Steps
Corresponding Graphic