GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-11 - Replacing Options
8 - 217
8-11-23-2
Removal Procedure
Steps
Corresponding Graphic
1.
Unscrew 10 screws to remove the Unity Rear Shield
Brkt from the console.
Note: If the WLAN-T was installed, remove the
the WLAN-T and WLAN USB A Cable" on page
2.
Unscrew 2 screws.
3.
Pull the UEPP Assy module to remove from the
console.
.
Summary of Contents for LOGIQ Totus