Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
8 - 237
8-12-78
Replacement of the UVSC-T
8-12-79
Replacement of the UBC-T
8-12-80
Replacement of the 4S2P Battery Pack (14.4V Li-ion 5900 mAh Rechargeable)
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
10-6-6 "Earth Leakage Current Test"
Attach the customer’s Internal Universal Video Converter
device to the rear I/O. Confirm content displayed on the
monitor mirrors content displayed on the main monitor of
the system. Confirm that the customer’s Internal Universal
Video Converter device can record content displayed on
the main monitor of the system.
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-25 "Battery Pack functional checks"
10-6-6 "Earth Leakage Current Test"
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-25 "Battery Pack functional checks"