Basic Service Manual
8 - 96
Section 8-7 - Replacing OPIO
8-7-8
Replacement of the Unity Gel Warmer Cable Cover
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-7-8-1
Removal Procedure
8-7-8-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-44 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
Philips Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Unscrew 1 screw to remove the Gel Warmer Cable Cover from the
console.