GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 131
8-8-3
Replacement of the Front Cover Assy-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-3-1
Preparations
1.) Remove the Unity Base Tray as described in
8-8-2 "Replacement of the Unity Base Tray" on page
8-8-3-2
Removal Procedure
8-8-3-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-65 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
No tools required
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1. Pull the Front Cover Assy-T to remove from the
console.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......