GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-11 - Replacing Options
8 - 189
8-11-6 Replacement of the Gel Warmer Cap
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-11-6-1
Removal Procedure
8-11-6-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-101 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
No tools required
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Twist the Gel Warmer Cap to remove it from the T2
Gel Warmer.