Basic Service Manual
Section 8-11 - Replacing Options
8 - 213
8-11-20 Replacement of the 3DG DriveBay 2+ EU and Power Adapter Cable
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-11-20-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
3.) Remove the VNAV Option Cable Set-U as described in
8-11-19 "Replacement of the VNAV Option
8-11-20-2
Removal Procedure
8-11-20-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-115 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Push the button of Power Adapter Cable to
disconnect Power Adapter Cable from the VNAV
Assy.