GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
8 - 214
Section 8-11 - Replacing Options
8-11-21 Replacement of the U-WIFI Module Assy
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-11-21-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the T2 Side Cover L Assy as described in
8-8-7 "Replacement of the T2 Side Cover L
8-11-21-2
Removal Procedure
8-11-21-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-116 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Unscrew 2 screws to remove the Cover bracket..
2.
Unscrew 1 screw of the WIFI module Assy and
disconnect the WIFI module from the UEPP
connector. Unscrew 4 screws to remove the WIFI
Module Assy from the console.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......