Basic Service Manual
8 - 222
Section 8-11 - Replacing Options
8-11-25 Replacement of the CWC-T PWA
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-11-25-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove Unity Base Tray as described in
8-8-2 "Replacement of the Unity Base Tray" on page 8-
3.) Remove Front Cover Assy-T as described in
8-8-3 "Replacement of the Front Cover Assy-T" on
4.) Remove Unity Protector Assy as described in
8-8-5 "Replacement of the Unity Protector Assy" on
8-11-25-2
Removal Procedure
8-11-25-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-120 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Unscrew 8 screws to remove the UAC&UCO-T Assy
2.
Unscrew 1 screw to remove the CWC-T PWA from
the UAC&UCO-T Module.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......