LOGIQ Totus
Basic Service Manual
2 - 2
Section 2-2 - General LOGIQ Totus Requirements
Section 2-2
General LOGIQ Totus Requirements
Contents in this section
2-2-1
LOGIQ Totus environmental requirements
If the LOGIQ Totus is very cold or hot
When unpacking the LOGIQ Totus, allow the temperature to stabilize before powering up.
Refer to
for guidelines on reaching operational temperatures from storage or transport
temperatures.
Environmental specifications
The LOGIQ Totus should be operated, stored, and transported according to the listed environmental
requirements.
NOTE:
Probe environmental specifications may be different than the LOGIQ Totus.
Refer to
for environmental specifications.
CAUTION
RISK OF DAMAGE TO EQUIPMENT
IF THE ULTRASOUND SYSTEM IS VERY COLD OR HOT, DO NOT TURN ON ITS POWER UNTIL
IT HAS HAD A CHANCE TO ACCLIMATE TO ITS OPERATING ENVIRONMENT.
Table 2-1 LOGIQ Totus Acclimate Time
o
C
50
45
40
35
30
25
20
15
10
5
0
-5
-10
o
F
122 113 104
95
86
77
68
59
50
41
32
23
14
Hrs
4
2
0
0
0
0
0
0
0
2
4
6
8
Table 2-2 Environmental Specifications for the LOGIQ Totus
Operational
Storage
Transport (< 16 hrs.)
Temperature
10° - 35°C
50° - 95°F
-10° - 50°C
14° - 122°F
-10° - 50°C
14° - 122°F
Humidity
30 - 80% non
-
condensing
10 - 90% non
-
condensing
10 - 90% non
-
condensing
Pressure
70 - 106kPa
70 - 106kPa
70 - 106kPa
Heat Dissipation
3070 Btu/hour
n/a
n/a
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......