GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
2 - 4
Section 2-2 - General LOGIQ Totus Requirements
2-2-2
Electrical requirements
General requirements
NOTE:
GE HealthCare requires a dedicated power and ground for the proper operation of its ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the LOGIQ Totus.
The LOGIQ Totus will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using 220
volt power, then a center tapped power source is required.
Sites with a mains power system with defined neutral and live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the LOGIQ Totus outlet.
Sites with a mains power system without a defined neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the LOGIQ Totus outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility’s incoming power source to the Ultrasound system is only a conduit.
Electrical requirements for LOGIQ Totus
, the electrical specifications for the LOGIQ Totus include monitor and on board peripherals.
Inrush current with UPS
During power on, an inrush circuit prevents the current from increasing above the stated values.
Refer to
for inrush currents at various main voltages.
Table 2-3 Electrical Specifications for the LOGIQ Totus
PART NUMBER
DESCRIPTION
VOLTAGE
TOLERANCES
POWER
CONSUMPTION
FREQUENCY
5926013/5943511 LOGIQ Totus, 100-240 VAC
100-240 VAC
+/-10%
0.65 kVA
50/60 Hz
Table 2-4 Inrush Current at Different Mains Voltages
VOLTAGE
Inrush Current
Remarks
115 VAC (Console only)
25 A Max
< 0.2 mSec
230 VAC (Console only)
50 A Max
< 0.2 mSec
120 VAC (Console with ABCE1442-11MED)
99 A Max
1~4 mSec
230 VAC (Console with ABCE1442-22MED)
99 A Max
1~4 mSec
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......