Lullaby
TM
Warmer 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 15
Over voltage and undervoltage indications.
Illustration
Description
Figure 21:
OVR
Figure 22:
UVR
If the applied mains voltage exceeds the
rated voltage by 10% for atleast 1
minute, “OVR” is displayed on the set
temperature display as seen in Figure 21.
The heating shall be cut off when “OVR” is
displayed on the display
WARNING
: Do not leave baby unattended
while “ovr” is displayed as heater will be
turned off.
If the applied voltage is lower than the
rated voltage by 10% for atleast 1
minute, “UVR” is displayed on the set
temperature display as seen in Figure 22.
In this condition, temperature monitoring
and all other warmer functions are
present.
Figure 23:
Heat OFF Indicator
Heat OFF indication:
When the hood is rotated to either of the sides
the heat level bar graph turns off and Heat
OFF indicator flashes.
NOTE: In the Heat OFF condition, audible
alarms are absent.
In this condition, temperature monitoring and
all other warmer functions are present.
Heat OFF
Indicator
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...