Revision D
MAC
1200 resting ECG analysis system
3-17
2012250-095
System Test Function: Setting the Device Model
The table below shows the most important combinations:
- Language international: g, e, f, i, s, por, sw, nor, fin, dan, hol, czech, hung
- Europe 2: United Kingdom, Benelux, Scandinavia
The required device model is selected by pressing the appropriate key,
1
to
6
.
Quit selection menu by pressing the
X
key.
Quit the self-test with
X
key.
MAC
1200 resting ECG analysis system configures the appropriate items and
initiates a cold start automatically. Thus, when the device is rebooted automatically
all the new settings are adopted.
Selecting a particular device model leads to the configuration of the program for
interpretation as indicated in the table above, even when the interpretation program
was configured differently beforehand as described in
“Interpretation (International
NOTE
When configuring the device model, saved ECGs are not lost.
If a new device model is selected, the program for interpretation is switched
over. When printing the saved ECGs, the display in the status line is related to
the currently enabled program for interpretation, although the results are based
on the program previously configured.
In this case, the saved ECGs should be printed out or transferred to a PC before
configuration of the new model.
Table 10. Model Combinations
Device Model
Interpretation
Output Formats
Languages
Default Settings
Override
Function Enabled
Line Filter
1 INT
HEART
international
international
no
50 Hz
2 USA
12SL
USA
e
yes
60 Hz
3 Europe 2
12SL
international
international
yes
50 Hz
4 Japan
12SL
international
e, jap
no
50 Hz
5 RUSS
HEART
international
e,r
no
50 Hz
6 Pharma
12SL
USA
e
no
60 Hz
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......