Maintenance
8.
Turn over the top cover assembly and secure the display assembly to the top
cover using the six fasteners shown in the following photograph.
The top cover assembly, with the display assembly mounted on it, is now
completely reassembled, as shown in the following photograph.
9.
Proceed to
“Reattaching the Top Cover Assembly to the Device” on page 81
Reattaching the Top Cover Assembly to the Device
Use the following procedure to reattach the top cover assembly to the device.
1.
Line up the top cover assembly to the device, as shown in the following
photograph.
2.
Gently lower the top cover assembly part way and connect the display cable (1)
and keypad ribbon (2) to the PWA, as shown in the following photograph.
2053535-003C
MAC™ 2000 ECG Analysis System
81