Troubleshooting: General Fault Isolation
MAC 5000 resting ECG analysis system
2000657-002
Power-up Flow Chart
Refer to the CPU PCB schematic diagram found in the PCB Assemblies
chapter at the end of the manual.
Poor Quality ECGs
Poor quality ECGs can be caused by factors in the environment,
inadequate patient preparation, hardware failures related to the
acquisition module, leadwires, cables, or problems in the unit.
Use a simulator to obtain an ECG report. If the report is good, the
problem is external to the unit.
Yes
Yes
Yes
Press ON key
Sh4
Micro-Controller
"MOE"
Sh8 pin 5
AC Present? Sh8
pin 13
Battery OK? Sh8
pin 17
Micro-Controller
"CURLY"
Sh2
configures FPGA
Sh2
"MOE" releases
nSysResest
Sh8 pin 10
Run application
software
Start-up screen
displayed
"CURLY" reads
SmartMedia card
and moves
instruction byte
to FPGA for
"unpacking"
StrongArm
access
SmartMedia card
via FPGA and
loads remaining
code to DRAM
"MOE" enables
+3VM and +5VM
four bytes loaded
to FPGA?
StrongArm
running?
Stabilization period
over?
"CURLY" asserts
WordRdy
Sh2 pin 9
FPGA puts
unpacked instruction
on StrongARM
data bus and
negates nWait
No
*
Reset
3VC
No
No
No
Yes
Yes
*
MD1322-017A
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...