REPLACEMENT OF THE PREFILTER, POSTFILTER, AND RO MEMBRANE ELEMENTS
17
Rev B
6. Remove exhausted prefilter and discard.
7. If Changing Membrane Elements:
A. Remove membrane sumps. Remove and discard used
elements.
B. Remove new elements from packaging.
C. Lubricate element O-rings, brine seals, and sump O-rings
with silicone lubricant.
D. Securely insert O-ring end of elements into manifold. See
Figure 8.
E. Replace sumps and hand-tighten.
8. Sanitize the system.
A. The manifold should be positioned flat with the sump
connections facing up.
Figure 8
CAUTION:
The person handling the filters and membrane elements
must have clean hands to keep the system sanitized. The use of
sterile/latex gloves is recommended.
CAUTION:
Elements contain a foodgrade preservative. The use of
sterile/latex gloves is highly recommended.
NOTE:
The system should be sanitized whenever a membrane
element or filter is replaced.
Prefilter Center Opening
Membrane Element Locations