2.9
Weights, Floor/Ceiling Loading and Recommended Mounting Methods
Table 2-5 Product Physical Characteristics (weight)
PRODUCT OR
COMPONENT
WEIGHT
LOAD BEARING
AREA ft2 (m2)
RECOMMENDED
MOUNTING
INFORMATION
Operator Console: PC
Tower HP z420 LCD
Monitor
19.6 Kg (43.2pounds)
8.2 kg (18.1 lbs)
Shelf or table
mounted but not
anchored.
Table Assembly
(Optima XR648 Only)
308.5 kg
(680lbs)
0.52 m2
(5.60 ft2)
Floor mounting
Recommendations:
(4) M10 X 120 mm
anchors (5727758,
supplied)
Table Assembly with a
maximum patient weight of
250kg (551lbs)
(Optima XR648 Only)
558.5kg
(1231lbs)
0.52 m2
(5.60 ft2)
NA
Stretchers (optional):
Non-elevating
Non-elevating (carbon fiber)
Elevating
102 kg (225 lbs)
32 kg (70.5 lb)
310 kg (683 lbs)
Not anchored
Not anchored
Not anchored
Wall Stand
(Standard global URP Wall
Stand)
270 kg (596 lbs)
0.129 m2
Bossard M10x150L
anchors to floor
(supplied)
Wall Stand
(Manual Wall Stand)
240 kg(530 lbs)
0.103 m2
Bossard M10x150L
anchors to floor
(supplied)
Extended Wall Stand
(Extended Global URP Wall
Stand)
280kg (618 lbs)
0.129 m2
Bossard M10x150L
anchors to floor
(supplied)
Stationary Rail (4 m)
48kg (106lbs) pair
OTS
240kg (572.6lbs)
2 Meter Bridge
54.8kg (120.8lbs)
3 Meter Bridge
63.7kg (140.43lbs)
Cable and Miscellaneous
Parts
53kg (117lbs)
Cable Drape Support
29.5kg (65lbs)
Upper OTS Covers
9.5kg (21lbs)
Stiffener Plate
10.4kg (23lbs)
UIF
3.2kg (7lbs)
System Cabinet
305kg (705lbs)
0.46 m2 (5 ft.2)
5/16 in. or 8 mm (6)
anchors to floor 5/
16 in. or 8 mm (6)
anchors to wall
(Mounting
hardware not
provided by GEHC)
Pre-Install.book Page 71 Tuesday, January 30, 2018 2:36 PM