Routine Maintenance
4.3
F-801 Cleaning Indications
Foulants such as minerals, silt and organic substances may collect on the RO membrane
surface and cause the RO membrane to foul and diminish in its function. Many of these
foulants are prevented by proper selection of pretreatment equipment, while others are
treatable only with more sophisticated designs. Regardless of the foulant, it is essential
to clean the membrane promptly once cleaning is indicated. The longer a foulant is
allowed to remain in contact with the RO membrane, the more difficult it will be to
remove. In severe cases, repeated cleanings may be required.
NOTE:
Some foulants CANNOT be removed with cleaning and membrane performance
may not be fully restored. Observe the appearance of the cleaning solution before
discarding it. If the solution is heavily soiled, repeat the cleaning procedure with
fresh solution of the same type (high pH or low pH).
Routine cleaning of the RO membranes in conjunction with regular disinfection will
keep bacteria counts at a minimum and the RO membrane performance at its best.
Cleaning removes the silt, organic and mineral deposits that collect on the RO
membrane surface which, if left untreated, encourage bacterial growth and will
eventually impair the RO membranes' ability to perform. Disinfection kills most of the
bacteria that may be present in the system, but it does not remove the deposits on the
membrane.
There are a variety of indications for cleaning membranes. The most common is a
reduced product flow rate. Remember that a reduction in inlet water temperature can
also result in a reduced flow rate. Refer to Technote 113 "Temperature Correction
Factors" to determine the reduction in flow that can be attributed to reduced water
temperature. Clean the membranes whenever the temperature compensated flow rate
drops by 10%. A loss in the membrane performance (percent rejection) or increase in
TDS of the product water may also be an indication to clean the membranes. A good
preventative maintenance practice is to clean the membranes on a quarterly basis, and
more often if needed.
CAUTION:
Membranes that have not been cleaned for a long time may actually incur damage
during cleaning as perforations are exposed that were masked by accumulations of
foulant. Regular maintenance cleaning is the best assurance of long membrane
service.
Summary of Contents for Osmonics F-801
Page 2: ......
Page 6: ...12552e D D 27 February 2004 This page intentionally left blank ...
Page 24: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 1 18 General Information NOTES ...
Page 39: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 3 7 7 System Operation NOTES ...
Page 73: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 4 33 Routine Maintenance NOTES ...
Page 87: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 5 13 Troubleshooting NOTES ...
Page 113: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 25 Drawings NOTES ...
Page 121: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 7 Spare Parts List NOTES ...
Page 123: ......