Chapter 4: Troubleshooting
Troubleshooting Tables
Service Manual
4-43
4.4.13 Mechanical (Bedded Models Only)
4.4.13.1 Bed Does Not Tilt
9.
Test fuses F2,F5. In addition, check the contacts on
the Warmer head harness. (Refer to
.) Replace/repair as necessary.
Warmer head
harness part #:
6600-1558-700
10.
If OK, reattach J28. There is AC voltage coming off
of the Power PCA but no DC being transferred
back.
11.
Otherwise, there is no AC voltage coming off of the
Power PCA. Proceed with the following steps.
12.
Disconnect connector at J27 on the Power PCA.
Check for AC power at pins 3,5. Also, test the front
panel switch (standby switch) using an Ohm Meter
at pins 1,2. Off= Open, On= Short. If Open, there is
no AC voltage to the Power PCA, then the front
panel switch is defective.
13.
If no power, troubleshoot the Line Filter Harness
from the AC inlet through the main circuit breaker,
through the line filter, through to the J27
connection (refer to
) to find
the open in the circuitry.
Possible Causes
Step
Troubleshooting Procedure
Comments
Possible Causes
Step
Troubleshooting Procedure
Comments
Disconnected
control cable
1.
Check that the cable terminators are fully seated in
retainers.
Bed lock
malfunctioning
2.
Isolate primary and secondary lock by
disconnecting one then the other from the main
brackets and then tilting the bed. Replace
defective lock as necessary.
Summary of Contents for PANDA
Page 4: ...RH 2 Service Manual...
Page 12: ...ii Service Manual About this Manual User Responsibility...
Page 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual...