Chapter 5: Repair Procedures
Electronic Enclosure Repairs
Service Manual
5-15
5.4.8 F7 and F8 Fuse Replacement Instructions
Fuse F7 and F8 are replaceable however the socket contacts are brass and extremely easy to bend causing
a loose connection when the new fuse is installed. Do not pry them out use needle nose pliers and pull
straight out. Reinsert gingerly the same way. Once damaged the socket can not be replaced in the field and
a new Power Bd will need to be installed. Refer to
for fuse ordering information.
5.4.9 Maneuvering Handle
Use a 13 mm wrench to remove the 2 nuts, split ring washers and flat washers from the mounting studs.
Slide the studs out the bottom of the dovetail rail.
F7
F8
Summary of Contents for PANDA