Appendix A: Specifications
Performance
A-2
Service Manual
A.4 Performance
System
Warmer Expected life
Approximately 8 years (see Life)
Heater Element
360 Watts
Heater Output
27 mW/cm
2
Patient temperature
measurement accuracy
± 0.3°C @ 30°C to 42°C
Temperature probe accuracy
± 0.1°C @ 30°C to 42°C
Observation Light
2 dimmable 35W halogen bulbs; estimated life 3000 hours based on
manufacturer’s specifications
Procedure Light
2000 lux (at nominal voltage) average; estimated life 3000 hours
Elevating base duty cycle
15%
Alarm Volume
At 3m away from the device, the alarm level is:
•
>60 dBA at the maximum setting
•
>50 dBA at the minimum setting
Life
The warmer is designed to last at least 8 years in normal use when operated, maintained, and serviced in
accordance with the instructions provided in the User Manual and Service Manual. With proper
maintenance and repairs, the service life can be extended as long as service parts are available.
Weight Scale
Functional range
300g to 8 kg
Accuracy
± 10g
Irradiance
50% Heater Power
100% Heater Power
Bedded Warmer
10-15 mW/cm
2
31 mW/cm
2
Highest bassinet
13 mW/cm
2
31 mW/cm
2
Summary of Contents for PANDA
Page 4: ...RH 2 Service Manual...
Page 12: ...ii Service Manual About this Manual User Responsibility...
Page 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual...