Appendix E: Installing Wall Mount Units
Enclosure Dimensional Specification
Service Manual
E-11
E.6 Enclosure Dimensional Specification
GE Healthcare does not offer an in-built bed, cabinet or other enclosure option for this product. If the facility
chooses to place the warmer in an enclosed space (e.g. cabinet, armoire), make sure that the enclosure
meets the minimum dimensional requirements as illustrated in
in order to have adequate air flow
to avoid warmer overheating when the enclosure is closed. During use (when the enclosure is open),
confirm that no object is blocking the radiant heat path to the bed (see
) and that the caregiver
can have unobstructed access to the warmer control panel, probe panel, rails and the bed surface. Do not
operate the warmer unless the cabinet, armoire or other enclosure is open (see
). GE Healthcare
will not install a warmer in an enclosure that does not provide the caregiver with unobstructed access to the
warmer control panel, probe panel, rails and the bed surface. The facility’s installation of, or operation of, the
warmer in such an enclosure is at the facility’s risk.
Warning:
Do not place objects in the radiant heat path. Objects will be heated and
could block heat to the baby.
Warning:
The caregiver shall have unobstructed access to the warmer control
panel, probe panel, dovetail rails and the bed surface during operation.
Summary of Contents for PANDA
Page 4: ...RH 2 Service Manual...
Page 12: ...ii Service Manual About this Manual User Responsibility...
Page 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual...