Chapter 4: Troubleshooting
Service Mode
Service Manual
4-3
4.1.4 Operation Settings
This menu enables users to customize the behavior of alarms, user functions, and displayed units by setting
default values for the parameters and options listed in the following table.
* The Low pulse rate limit can be set only between 25 and the High pulse rate limit -1.
The High pulse rate limit can be set only between the Low pulse rate limit +1 and 240.
Feature
Available Settings
US Defaults
International Defaults
Patient temp. alarm
0.5°C / 1.0°C
1.0°C
1.0°C
Check patient alarm disabled
5% - 50%
25%
25%
Man. temp. alarm
Off / 0.5°C / 1.0°C
Off
Off
Alarm volume
0 / 1 / 2 / 3
3
3
Hands free silence
Enable / Disable /
Disengaged
Enable
Enable
Bed elevate (for Bedded Warmers
with Elevating Base)
Enable / Disable /
Disengaged
Enable
Enable
Bed elevate (for Bedded Warmers
with Fixed-Adjustable Base,
Freestanding Warmers, and Wall
Mount Warmers)
Enable / Disable /
Disengaged
Disengaged
Disengaged
Temp. units
C / C Only / F
C
C Only
Scale units
grams / lbs-oz /
grams only
grams
grams only
Scale resolution
10/5 grams
10 grams
10 grams
Language
Many
English
English
Low pulse rate limit*
25 - 110
100
100
High pulse rate limit*
111 - 240
200
200
Low saturation limit**
20 - 99
85
85
High saturation limit**
86 - 100
100
100
SpO
2
alarm suspension on startup
0 / 15 / 30
15
15
SpO
2
averaging (Masimo only)
2-4 / 8 / 16
8
8
SpO
2
sensitivity (Masimo only)
Normal / Maximum
Normal
Normal
SpO
2
SatSeconds (Nellcor only)
Off, 10, 25, 50, 100
Off
Off
SpO
2
Response Mode (Nellcor only)
Normal / Fast
Normal
Normal
SpO
2
Display (Nellcor only)
Pleth / Blip
Pleth
Pleth
SpO
2
View
Large / Normal
Large
Large
Pulse tone volume, Panda
0/ 1 / 2 / 3 / 4
1
1
Time Format
0/ 1 / 2 / 3 / 4
12
24
Date Format
YYYY/MM/DD
MM/DD/YYYY
DD/MM/YYYY
MM/DD/YYYY
DD/MM/YYYY
APGAR TONES (Giraffe Only)
ON/OFF
OFF
OFF
Summary of Contents for PANDA
Page 4: ...RH 2 Service Manual...
Page 12: ...ii Service Manual About this Manual User Responsibility...
Page 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual...