Appendix C: Alarm Priorities
Priority Alarm Subject
Activation
Criteria
Message
Displayed
Silence
Time
Sound:
1 = High
2 = Low
3 = informational
Notes/
Comments
Indication
Light
Special
Requirements
1
SpO
2
lost
communica-
tions
SpO
2
board
has not
communicated
with the main
processor for
at least
2 seconds
Oximetry
System
Failure
Refer
to
notes
1
When this alarm
occurs, the SpO
2
module will be
automatically
turned off and
SpO
2
options will
no longer be
available. When
silenced, the
alarm will clear.
Primary
2
Pulse rate
PR < 100 (floats
with user
adjustable
limits)
Low Pulse
Rate
3 min.
1
Primary
3
Oxygen
saturation
SpO
2
<85
(floats with
user adjustable
limits)
Low SpO
2
3 min.
1
Primary
4
Check baby
Radiant %
power @ 100%
for >15 min
and alarm not
silenced
Check
Baby -
Heat Off
15 min
1
Primary
Operational in
Baby Control
mode
5
Check baby
Radiant heater
% above
“Check Baby
Alarm
Disabled”
(CPAD) limit for
>15 min and
alarm not
silenced
Check
Baby -
Heat Off
15 min
1
Primary
Operational in
Baby Control
Mode
6
Baby
temperature
>2.0° from
Control Temp
Baby Hot-
Check
Temp
Probe
5 min
1
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 1.8°C
Primary
Operational in
Baby Control
Mode
Summary of Contents for PANDA
Page 4: ...RH 2 Service Manual...
Page 12: ...ii Service Manual About this Manual User Responsibility...
Page 38: ...Chapter 1 Functional Description System Functions 1 14 Service Manual FIGURE 1 8 Power Failure...
Page 60: ...Chapter 2 Installation Maintenance and Checkout Scale Checkout Procedures 2 8 Service Manual...
Page 62: ...Chapter 3 Calibration Scale Calibration Bedded Models Only 3 2 Service Manual...
Page 172: ...Chapter 6 Illustrated Parts Wiring Diagrams 6 40 Service Manual...
Page 180: ...Appendix A Specifications SpO2 Specifications A 8 Service Manual...