Performance RHM 15
Standard Models
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
200
441
0.20
100
220
0.20
50
110
0.20
10
22
0.20
4
8.8
0.50
Goldline Models (**) - selected sensors
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
150
331
0.10
75
165
0.10
50
110
0.10
15
33.1
0.10
7.5
16.5
0.12
Low Flow Models (**) - selected sensors
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
80
176
0.20
50
110
0.20
15
33
0.20
4
8.8
0.20
2
4.4
0.60
Repeatability
Better ± 0.1% of rate,
Goldline 0.05%
Density
Down to 0.5% uncertainty
Temperature
Better ± 1°C
(*) Nominal flow Qnom refers to roughly 8 - 10
m/s (26 - 33 ft/s) velocity in measuring loops
for best performance.
(**) Selected sensors are only available in
combination with temperature options T1, TA,
standard material and pressure range.
• Serial/single path versions offer the same accuracy at
half the flow (Qmax serial version = 100 kg/min).
• No relevant pressure effect due to torsional oscillation
and semi-circle (non-deforming) measurement
section.
• Uncertainty of reading (incl. zero drift) indications refer
to reference conditions H
2
O, 18-24°C (66 - 76°F), 1 - 3
bar (15 – 45 psi) and installation according to field
manual.
• Pressure drop indications refer to H
2
O, with parallel
measuring loops type P1 and standard manifold block
connections.
• For calibration to customer range and / or with
improved uncertainty, please consult factory.
Uncertainty % DP (bar)
Uncertainty % DP (bar)
Uncertainty % DP (bar)
Flow Rate kg/min
Flow Rate kg/min
Flow Rate kg/min
Max Flow Rate Qmax = 200 kg/min (441 lb/min), Qnom (*) = 150 kg/min (331 lb/min)
6