GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
124
2.
PRINTER MANAGEMENT
Printers are set up and declared to the system at the time of installation by the GE Field Engineer;
printer information cannot be changed by the user.
The access to this function is from the tools menu
3.
PRINT FUNCTIONS
3-1
Auto Print
Auto Print is a function allowing the user to automatically initiate the printing of exam images on
closure of the exam from the AWS.
To use Auto Print, access the Auto Print setup tool. Click on
Auto Print
in the
Medical Application
preferences
window (available in the Browser
Tools menu
menu, as described in section 8 of
Chapter 7,
Browser
).
Only images with a Quality Check value of OK are automatically printed by the Auto Print function.
Images with a Quality Check value of
NOT OK
are not automatically printed at the close of the exam.
Facilities available from the
Auto Print
window are:
D
DICOM printers
: contains the list of all the available DICOM printers that have been declared
for printing. Click on a printer in the list to select it.
D
Auto Print
buttons: click the On
button to turn Auto Print on for the selected printer or the Off
button to turn it off. The selected button becomes dark gray.
D
Save
button: click this button to confirm any changes made and close the window.
D
Close
button: click this button to close the
Medical Application preferences
window. If the user
has made changes to the Auto Print window, a Warning window opens requesting the user to
either use:
– Save button (to save the changes and return to the Browser).
– Cancel button (to close the Warning window and return to the Auto Print window).
– Discard button (to close without saving the changes and return to the Browser).
D
Print mode: only Fit to Film. Can not be changed by the user in current Senographe 2000 D
release.
D
Number of Copies
box: enter the required number of copies in this box.
D
Printing properties
button: click this button to open the
Printing Properties
window
(see below § 3-3).