GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A009 - AWS Bootup Troubleshooting
Page no. 1017
Chapter 9
JC-TSG-A-009.fm
Job Card TSG A009 - AWS Bootup Troubleshooting
Chapter 9
1
INTRODUCTION
This Job Card is intended as an aid to troubleshooting Acquisition Workstation (AWS) bootup problems.
It is used when a bootup fault prevents the AWS booting up correctly.
1-1
Blank Screen after Logging in as the sdc User
Once someone has logged in at the first AWS
Login:
prompt with the
sdc
user, the monitor goes blank
(white with no text or graphics) and then returns to the
Login:
prompt. This scenario occurs in the follow-
ing situations:
•
the mouse or trackball is disconnected
•
the mouse or trackball is faulty
If you witness a blank screen that keeps returning to the
Login:
prompt after logging in as the
sdc
user
then you must do the following:
•
Check that the mouse/trackball is correctly attached to the AWS. Check both the connections on the
back panel of the ADS computer and on the USB extension cable that connects the mouse/trackball
to the ADS computer.
•
If the problem persists after you have check the cable connections, the mouse/trackball is faulty. You
must change the mouse/trackball according to
or